Get RTF file here and follow the instructions. Depending on your configuration, a right click on that link might open it or might produce a "do you want to save" message. Save it to a local or network drive and keep track of it. If your file extensions are showing and you don't change the name, it is called BRWN_WTHR_2002.rtf Any recent word processor version on a PC or Mac should open it.
|
bldg |
room |
responsible |
Item of concern |
abatement time |
date corrected |
| BRWN | B125A | Wenthold | Workplace not reasonably free of recognized hazards. Compressed gas cylinder containing 5% fluorine:95% helium not in vented gas cabinet. | 60 days | |
| BRWN | B137 | Ben-Amotz | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Specific PPE requirements needed instead of generic. | 5 days | 1/13/2003 |
| BRWN | B165A | Francisco(?) | Unsecured or inadequately secured compressed gas cylinder(s). Uncapped cylinders without regulators are unacceptable. Cylinders with regulators should be secured individually. Small cylinder not secured. | Immed | |
| BRWN | B178 | McLuckey | Flexible cords and/or cables not protected from accidental damage. Power cord across walkway needs to be removed. Cord tunnel not acceptable. | Immed | 1/17/2003 |
| BRWN | Basement Corridor |
Dirksen | Corridor is obstructed by 42 pallets of office partitions and assembly parts. Items need to be removed from the corridor. | 60 days | 1/16/2003 |
| BRWN | 1142 | Swihart | Workplace not reasonably free of recognized hazards: duplex outlet in kitchenette near sink is without GFCI protection. Submit OSHA 113 to REM. REM will issue 18-A. P59 is not required. | 60 days | 113 submitted 1/24/2003 |
| BRWN | 1152B | Restore aisleway. Grocery sacks of material, carpeting roll, cloth, and board obstructs the aisleway. | 10 days | ||
| BRWN | 2105 | Restore aisleway. | 10 days | ||
| BRWN | 2105 | Heat resistant pad (ceramic tile or metal stove and counter top pad) needed under coffee maker(s) and/or hot water heater(s). (Needed for large coffee maker) | 60 days | ||
| BRWN | 2105 | Electrical cord tunnels across aisleways are not acceptable. | 30 days | ||
| BRWN | 2107B | Oakley | Heat resistant pad (ceramic tile or metal stove and counter top pad) needed under coffee maker(s) and/or hot water heater(s). | 60 days | 12/12/02 |
| BRWN | 2184 | Keyes | Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC | 10 days | 12/12/2002 |
| BRWN | 2184 | Keyes | Display written SOP for accessing extremely high containers of liquids, or move containers of liquids stored above eye level to eye level or below. | 30 days | 12/13/2002 |
| BRWN | 2184 | Keyes | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. | 5 days | 12/12//2003 |
| BRWN | 3124 | Hrycyna | Eyewash and/or shower is obstructed. | Immed | 1/9/2003 |
| BRWN | 3124 | Hrycyna | Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls | Immed | 1/9/2003 |
| BRWN | 3124 | Hrycyna | Unsecured or inadequately secured compressed gas cylinder(s). Uncapped cylinders without regulators are unacceptabl. (3 cylinders without regulators ganged together.) | Immed | 1/9/2003 |
| BRWN | 3124 | Hrycyna | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific. | 5 days | 1/9/2003 |
| BRWN | 3124 | Hrycyna | Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC | 10 days | 1/9/2003 |
| BRWN | 3124A | Laskowski | Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC | 10 days | 12/20/2002 |
| BRWN | 3124A | Laskowski | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific. | 5 days | 12/20/2002 |
| BRWN | 3124A | Laskowski | Failure to designate areas for select carcinogens, reproductive hazards, and/or extremely toxic chemicals. Use supplied "CAUTION" stickers as needed in these areas. See http://www.purdue.edu/rem/home/booklets/crdalist.doc (Chloroform) | 10 days | 12/20/2002 |
| BRWN | 3124B | Hrycyna | Eyewash and/or shower is obstructed. | Immed | 1/9/2003 |
| BRWN | 3124B | Hrycyna | Excessive fire fuel loading (paper and/or other combustibles). Excessive accumulation of styrofoam containers. | Immed | 1/9/2003 |
| BRWN | 3124B | Hrycyna | Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls | Immed | 1/9/2003 |
| BRWN | 3124B | Hrycyna | Display written SOP for accessing extremely high containers of liquids, or move containers of liquids stored above eye level to eye level or below. | 30 days | 1/9/2003 |
| BRWN | 3124B | Hrycyna | Open chemical containers not in use. (5 gallon carboy uncapped.) | Immed | 1/9/2003 |
| BRWN | 3124B | Hrycyna | Container of chemical is unlabeled. (5 gallon carboy unlabeled.) | Immed | 1/9/2003 |
| BRWN | 3124B | Hrycyna | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific. | 5 days | 1/9/2003 |
| BRWN | 3124B | Hrycyna | Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC | 10 days | 1/9/2003 |
| BRWN | 3125A | Chmielewski | Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC | 10 days | 12/17/2002 |
| BRWN | 3125A | Chmielewski | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. | 5 days | 12/16/2002 |
| BRWN | 3125B | Chmielewski | Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls | Immed | 12/17/2002 |
| BRWN | 3134 | Laskowski | Container of chemical is unlabeled. | Immed | 12/20/2002 |
| BRWN | 3134 | Laskowski | Combustible materials not stored safely. (Excessive saw dust accumulation in fume hood B.) | Immed | 12/20/2002 |
| BRWN | 3134 | Laskowski | Chemical hood, glovebox, and/or laminar flow filtered air unit unsafe due to excessive accumulation of containers and other materials. Flammables stored next to an ignition source in fume hood A. | 10 days | 12/20/2002 |
| BRWN | 3134A | Laskowski | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. | 5 days | 12/20/2002 |
| BRWN | 3134A | Laskowski | Food stored with laboratory equipment and/or chemicals in walk-in. | Immed | 12/13/2002 |
| BRWN | 3134B | Laskowski | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific. | 5 days | 12/20/2002 |
| BRWN | 3134C | Laskowski | Display written SOP for accessing extremely high containers of liquids, or move containers of liquids stored above eye level to eye level or below. | 30 days | 12/20/2002 |
| BRWN | 3134C | Laskowski | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific. | 5 days | 12/20/2002 |
| BRWN | 3134C | Laskowski | Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC | 10 days | 12/20/2002 |
| BRWN | 3134E | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific. | 5 days | ||
| BRWN | 3135 | Container of chemical is unlabeled. (Carboy unlabeled.) | Immed | ||
| BRWN | 3135A | Eyewash and/or shower is obstructed. | Immed | ||
| BRWN | 3135E | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. | 5 days | ||
| BRWN | 3144 | Laskowski | Container of chemical is unlabeled. | Immed | 12/20/2002 |
| BRWN | 3144 | Laskowski | Restore aisleway. | 10 days | 12/20/2002 |
| BRWN | 3144 | Laskowski | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific. | 5 days | 12/20/2002 |
| BRWN | 3151A corridor | Corridor obstructed with bookcases and filing cabinet. | 10 days | ||
| BRWN | 3151E | Eyewash and/or shower is obstructed. Eyewash is obstructed by curtain on safety shower. | Immed | ||
| BRWN | 3151E | Electrical cord tunnels across aisleways are not acceptable. | 30 days | ||
| BRWN | 3151E | Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC | 10 days | ||
| BRWN | 3151F | Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC | 10 days | ||
| BRWN | 3151F | Chemical hood, glovebox, and/or laminar flow filtered air unit unsafe due to excessive accumulation of containers and other materials. Flammables stored next to an ignition source in both fume hoods. | 10 days | ||
| BRWN | 3164 | Eyewash and/or shower is obstructed. | Immed | ||
| BRWN | 3164 | Container of chemical is unlabeled. | Immed | ||
| BRWN | 3164 | Display written SOP for accessing extremely high containers of liquids, or move containers of liquids stored above eye level to eye level or below. | 30 days | ||
| BRWN | 3164 | Failure to designate flam. storage cabinets for select carcinogens, reproductive hazards, and/or extremely toxic chemicals. Use supplied "CAUTION" stickers as needed in these areas. See http://www.purdue.edu/rem/home/booklets/crdalist.doc | 10 days | ||
| BRWN | 3164 | Unsecured or inadequately secured compressed gas cylinder(s). Uncapped cylinders without regulators are unacceptable. Cylinders with regulators should be secured individually. (Improperly secured helium cylinder.) | Immed | ||
| BRWN | 3164 | Electrical equipment shall be free from recognized hazards that are likely to cause death or serious physical harm to employees. (Hot plate is missing the base plate.) | Immed | ||
| BRWN | 3170 | Unsecured or inadequately secured compressed gas cylinder(s). Uncapped cylinders without regulators are unacceptable. Cylinders with regulators should be secured individually. (4 unsecured cylinders.) | Immed | ||
| BRWN | 3171 | Container of chemical is unlabeled. | Immed | ||
| BRWN | 3171 | Unsecured or inadequately secured compressed gas cylinder(s). Uncapped cylinders without regulators are unacceptable. Cylinders with regulators should be secured individually. | Immed | ||
| BRWN | 3174 | Clean broken glass or plastic sharps should be collected in a cardboard box or other strong, secure disposable container. Needles, razor blades, scalpels, etc., must be disposed of in puncture proof plastic containers. Sharps container was over filled. | Immed | ||
| BRWN | 3174 | Restore aisleway. (CORRECTED by REM.) | 10 days | ||
| BRWN | 3174 | Display written SOP for accessing extremely high containers of liquids, or move containers of liquids stored above eye level to eye level or below. | 30 days | ||
| BRWN | 3174 | Chemical containers have labels that are deteriorating and/or unreadable in fume hood A. | Immed | ||
| BRWN | 3175 | Keyes | Power cord damaged or broken. Electrical tape is unacceptable. (End of electrical cord on Thelco brand lab equipment is broken.) | 10 days | 12/13/2002 |
| BRWN | 3175 | Keyes | Container of chemical is unlabeled. | Immed | 12/13/2002 |
| BRWN | 3182 | Display written SOP for accessing extremely high containers of liquids, or move containers of liquids stored above eye level to eye level or below. | 30 days | 12/12/2002 | |
| BRWN | 5125 | Ramachandran | Open chemical containers not in use. | Immed | 12/12/2002 |
| BRWN | 5125 | Ramachandran | Excess flammables stored outside of approved cabinet and/or safety cans. There are flam. cabs used for non flam. storage and excess flam. liquids outside of appropriate storage. Only 10 gal total Class I plus Class II liquids outside ot flam. storage. | 60 days | 12/12/2002 |
| BRWN | 5125 | Ramachandran | Flammable storage cabinet not closed properly. Must be kept closed to be effective. Corrected at inspection. | Immed | 10/28/2002 |
| BRWN | 5125 | Ramachandran | Excessive use of floor space for containers of chemicals, equipment, and/or other items restrict aisleways. Remove ALL chemicals and waste from walking and working aisles, also equipment, also cans of used syringes and needles, also safety cans. | Immed | 12/15/2002 |
| BRWN | 5125 | Ramachandran | Excessive accumulation or abandoned containers of chemical waste. | 10 days | 12/13/2002 |
| BRWN | 5125 | Ramachandran | Write date on waste containers at start of accumulation. Waste is very poorly managed in this room and this step is required that more attention be paid to it. | Immed | 12/16/2002 |
| BRWN | 5125 | Ramachandran | Employee observed working with liquids without splash goggles, required by sign on hallway side of 5125 sign. | Immed | 12/12/2002 |
| BRWN | 5125 | Ramachandran | Hazard assessment certification is posted but is not appropriate to this work area. Remove WELDING and any other sections not necessary. Ensure that rules are clear, appropriate and enforced. | 10 days | 1/16/2003 |
| BRWN | 5125 | Ramachandran | No employee shall be allowed to consume food or beverages … in any area exposed to a toxic material. This room shall be posted completely off-limits for all food and beverage. | Immed | 12/16/2002 |
| BRWN | 5125 | Ramachandran | Hoods function properly when work is set up 6" from face opening. Add a tape or paint line to any/all hoods where necessary to remind users to keep the work 6" back. | 60 days | 12/12/2002 |
| BRWN | 5129 | Ramachandran | Unsecured compressed gas cylinder(s), also other cylinders ganged behind a single support chain. Cylinders should be secured individually while cap is off. | Immed | |
| BRWN | 5131 | Ramachandran | Failure to use required personal protective equipment. Main entance sign requires wearing of goggles inside at all times. Remove sign or enforce. | Immed | 12/12/2002 |
| BRWN | 5134 | Rothwell | Excess flammables stored outside of approved cabinet and/or safety cans. A flam cab exists which contains many NON flammables. Take out nonflammables, put in flammables. | 60 days | |
| BRWN | 5134 | Rothwell | Hazard assessment certification is posted but is not appropriate to this work area. Remove WELDING and any other sections not necessary. Ensure that rules are clear, appropriate and enforced. | 10 days | |
| BRWN | 5134A | Rothwell | Open chemical waste containers not in use. Funnel top cover was in up position. Corrected at inspection. | Immed | |
| BRWN | 5135 | Ramachandran | Excessive use of floor space for containers of chemicals, equipment, and/or other items restrict aisleways. Remove ALL chemicals and waste from walking and working aisles, also equipment, also cans of used syringes and needles, also safety cans. | Immed | 12/13/2002 |
| BRWN | 5135 | Ramachandran | Write date on waste containers at start of accumulation. Waste is very poorly managed in this room and this step is required that more attention be paid to it. | Immed | 12/14/2002 |
| BRWN | 5135 | Ramachandran | Hoods function properly when work is set up 6" from face opening. Add a tape or paint line to any/all hoods where necessary to remind users to keep the work 6" back from face of hood. | 60 days | 12/12/2002 |
| BRWN | 5135 | Ramachandran | No employee shall be allowed to consume food or beverages … in any area exposed to a toxic material. This room shall be posted completely off-limits for all food and beverage. | Immed | 12/12/2002 |
| BRWN | 5135 | Ramachandran | Excessive accumulation or abandoned containers of chemical waste. Nearly 50 gallons of waste in over 15 containers, mostly on the floor. Waste MUST be managed more effectively. | 10 days | 12/13/2002 |
| BRWN | 5135 | Ramachandran | Excess flammables stored outside of approved cabinet and/or safety cans. Flam cab in this room is extremely under-utilized, could take more flammables if shelves were replaced. | 60 days | 12/12/2002 |
| BRWN | 5135 | Ramachandran | Hazard assessment certification is posted but is not appropriate to this work area. Remove WELDING and any other sections not necessary. Ensure that rules are clear, appropriate and enforced. | 10 days | 1/16/200 |
| 3BRWN | 5135 | Ramachandran | Failure to use personal protective equipment required by hazard assessment certification. Employee observed pouring liquids without splash goggles. | Immed | 12/12/2002 |
| BRWN | 5150 | Fenniri | Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls | Immed | |
| BRWN | 5150 | Fenniri | Flammable storage cabinet not closed properly. Must be kept closed to be effective. Corrected at inspection. | variable | |
| BRWN | 5150 | Fenniri | Hoods function properly when work is set up 6" from face opening. Add a tape or paint line to any/all hoods where necessary to remind users to keep the work 6" back. | 60 days | |
| BRWN | 5150 | Fenniri | Excess flammables stored outside of approved cabinet and/or safety cans. | 60 days | |
| BRWN | 5150 | Excessive use of floor space for containers of chemicals, equipment, and/or other items restrict aisleways. | Immed | ||
| WTHR | 011, Stairwell nearby | Fire doors must not be left propped open to this stairwell. | Immed | ||
| WTHR | 012 | Dirksen | Restore aisleways. Aisleway blocked on the south side and the east side of the room. | 10 days | 1/16/2003 |
| WTHR | 031 | Cooks | Workplace not reasonably free of recognized hazards: duplex outlet in kitchenette without GFCI protection. Submit OSHA 113 to REM. REM will issue 18-A. P59 is not required. | 60 days | 113 submitted 1/24/2003 |
| WTHR | 033C | Cooks | Fire doors must not be left propped open when room is unattended. Twin firedoors chocked open to 033C. | Immed | 12/16/2002 |
| WTHR | 033C | Cooks | Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC | 10 days | 12/18/2002 |
| WTHR | 033D | Cooks | Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC | 10 days | 12/18/2002 |
| WTHR | 040 | No backflow preventers or flashback arresters installed on oxygen and acetylene tank of welding outfit. One of each needed at each tank outlet. | 60 days | ||
| WTHR | 055A | Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC | 10 days | ||
| WTHR | 060 | Container of chemical is unlabeled. Five gallon carboy on west shelf containing a clear liquid is without contents and hazard labels. | Immed | ||
| WTHR | 062 | Cooks | Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls | Immed | 12/18/2002 |
| WTHR | 062 | Cooks | Open chemical containers not in use. Funnel in 5 gallon methylene chloride open container. | Immed | 12/17/2002 |
| WTHR | 062 | Cooks | Flexible cords and/or cables not protected from accidental damage. Computer cable across walkway at NE door to room needs to be removed. | Immed | 1/03/2003 |
| WTHR | 070 | Container of chemical is unlabeled. Spray bottle with blue liquid is without label of contents and hazards. | Immed | ||
| WTHR | 075 | Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC | 10 days | ||
| WTHR | 077 | Cooks | Unsecured or inadequately secured compressed gas cylinder(s). Uncapped cylinders without regulators are unacceptable. Cylinders with regulators should be secured individually. | Immed | 12/18/2002 |
| WTHR | 081 | Container of chemical is unlabeled. Squeeze bottle with clear liquid is without contents and hazard labels. | Immed | ||
| WTHR | 081 | Food stored with laboratory equipment and/or chemicals. Food stored in chemical refrigerator with chemicals. | Immed | ||
| WTHR | 083, in corridor | Dirksen | Corridor obstructed by 13 desks and 7 benches between elevator and room 083. | 10 days | 1/17/2003 |
| WTHR | 101 | Flexible cords and/or cables not protected from accidental damage. Tubing across walkway needs to be removed. | Immed | ||
| WTHR | 105 | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. | 5 days | ||
| WTHR | 110B | Lipschutz | Failure to designate areas for select carcinogens, reproductive hazards, and/or extremely toxic chemicals. Use supplied "CAUTION" stickers as needed in these areas. See http://www.purdue.edu/rem/home/booklets/crdalist.doc ( Formaldehyde) | 10 days | |
| WTHR | 110B | Lipschutz | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Specific not generic assessment needed. | 5 days | |
| WTHR | 117A | Nash | Heat resistant pad (ceramic tile or metal stove and counter top pad) needed under coffee maker.* | 60 days | 12/16/2002 |
| WTHR | 120A | Interference to sprinkler discharge because of building contents. Minimum vertical clearance between sprinklers and material below shall be 18": cardboard boxes on the top row. | Immed | ||
| WTHR | 127 | Combustible materials not stored safely. More than 100 empty cardboard boxes stored in north central part of room need to be collapsed for storage. | Immed | ||
| WTHR | 141 (140A*) | Dirksen | Grinder's work rest is greater than 0.125" from abrasive wheel. One wheel is missing tool rest on 1/2 H.P. Sears bench grinder. | 10 days | 1/17/2003 |
| WTHR | 144 | Rothwell/Wood | Unsecured or inadequately secured compressed gas cylinder(s). Uncapped cylinders without regulators are unacceptable. Cylinders with regulators should be secured individually. | Immed | 12/13/2002 |
| WTHR | 144 | Rothwell/Wood | Container of chemical is unlabeled. Squeeze bottle on sink is without contents and hazard labels. | Immed | 12/13/2002 |
| WTHR | 165 | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Specific not generic PPE assessment needed. | 5 days | 12/2/2002
|
|
| WTHR | 168 | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Specific not generic PPE assessment needed. | 5 days | 12/2/2002
|
|
| WTHR | 173B | Paul Bower | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Specific not generic PPE assessment needed. | 5 days | 12/2/2002 |
| WTHR | 173B | Paul Bower | Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC | 10 days | 12/20/2002 |
| WTHR | 175 | Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC | 10 days | 1/9/03 | |
| WTHR | 175 | Unsecured or inadequately secured compressed gas cylinder(s). Uncapped cylinders without regulators are unacceptable. Cylinders with regulators should be secured individually. At west end of room, 2 cylinders inadequately secured. | Immed | 12/2/2002 | |
| WTHR | 175 | Failure to designate areas for select carcinogens, reproductive hazards, and/or extremely toxic chemicals. Use supplied "CAUTION" stickers as needed in these areas. See http://www.purdue.edu/rem/home/booklets/crdalist.doc (Toluene) | 10 days | 12/20/2002 | |
| WTHR | 175 | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. | 5 days | 12/20/2002 | |
| WTHR | 185A | Restore aisleway. East and west door aisles obstructed. | 10 days | ||
| WTHR | 206 | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific. | 5 days | ||
| WTHR | 206 | Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC | 10 days | ||
| WTHR | 206 | Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls | Immed | ||
| WTHR | 206 | Container of chemical is unlabeled. Unlabeled bottles in fume hoods A & B. | Immed | ||
| WTHR | 208 | Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls | Immed | ||
| WTHR | 208 | Failure to designate areas for select carcinogens, reproductive hazards, and/or extremely toxic chemicals. Use supplied "CAUTION" stickers as needed in these areas. See http://www.purdue.edu/rem/home/booklets/crdalist.doc | 10 days | ||
| WTHR | 208 | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific. | 5 days | ||
| WTHR | 208 | Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC | 10 days | ||
| WTHR | 217 | Heat resistant pad (ceramic tile or metal stove and counter top pad) needed under coffee maker(s) and/or hot water heater(s).* | 60 days | ||
| WTHR | 221 | Heat resistant pad (ceramic tile or metal stove and counter top pad) needed under coffee maker(s) and/or hot water heater(s).* | 60 days | ||
| WTHR | 221B | Heat resistant pad (ceramic tile or metal stove and counter top pad) needed under coffee maker(s) and/or hot water heater(s).* | 60 days | ||
| WTHR | 225 | Electrical cord tunnels across aisleways are not acceptable. | 30 days | 1/31/2003 | |
| WTHR | 227A | Heat resistant pad (ceramic tile or metal stove and counter top pad) needed under coffee maker(s) and/or hot water heater(s).* | 60 days | 1/31/2003 | |
| WTHR | 236 | Required exit(s) obstructed. Exit to hallway is obstructed. | Immed | ||
| WTHR | 242 | Container of chemical is unlabeled. | Immed | ||
| WTHR | 242 | Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls | Immed | ||
| WTHR | 242 | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. | 5 days | ||
| WTHR | 248 | Container of chemical is unlabeled in fume hood A. | Immed | ||
| WTHR | 248 | Open chemical containers not in use in fume hood A. | Immed | ||
| WTHR | 250A | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. | 5 days | ||
| WTHR | 250A | Container of chemical is unlabeled. | Immed | ||
| WTHR | 250B | Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls | Immed | ||
| WTHR | 265 | Heat resistant pad (ceramic tile or metal stove and counter top pad) needed under coffee maker(s) and/or hot water heater(s). (3 needed)* | 60 days | ||
| WTHR | 316 | Container of chemical is unlabeled. (Unlabeled container on benchtop.) | Immed | ||
| WTHR | 317 | Container of chemical is unlabeled. (Unlabeled container on benchtop.) | Immed | ||
| WTHR | 327 | Low | No employee shall be allowed to consume food or beverages in a toilet room nor in any area exposed to a toxic material. Evidence of drinking in Biohazard designated area. | Immed | 1/21/2002 |
| WTHR | 327 | Low | Clean broken glass or plastic sharps should be collected in a cardboard box or other strong, secure disposable container. Needles, razor blades, scalpels, etc., must be disposed of in puncture proof plastic containers. Sharps container was over filled. | Immed | 1/17/2003 |
| WTHR | 327 | Low | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific and clean copy. | 5 days | 1/21/2003 |
| WTHR | 327 | Low | Unsecured or inadequately secured compressed gas cylinder(s). Uncapped cylinders without regulators are unacceptable. Cylinders with regulators should be secured individually. (Carbon dioxide cylinder was without a cap or regulator.) Corrected at inspection. | Immed | 10/28/2002 |
| WTHR | 327 | Low | Container of chemical is unlabeled. (In Biosafety Cabinet.) | Immed | 1/17/2003 |
| WTHR | 327 | Low | Open chemical containers not in use. (In Biosafety Cabinet.) | Immed | 1/17/2003 |
| WTHR | 327 | Low | Display written SOP for accessing extremely high containers of liquids, or move containers of liquids stored above eye level to eye level or below. | 30 days | 1/17/2003 |
| WTHR | 327 | Low | Stored chemicals are not segregated by hazards. (Hood A - perchloric acid stored with acetic acid.) | Immed | |
| WTHR | 327 | Low | Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls | Immed | |
| WTHR | 327B | Low | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. | 5 days | 1/21/2003 |
| WTHR | 327C | Low | Heat resistant pad (ceramic tile or metal stove and counter top pad) needed under coffee maker(s) and/or hot water heater(s).* | 60 days | Kevin fixed 331C, not this one |
| WTHR | 327E | Low | Open chemical containers not in use. | Immed | 1/17/2003 |
| WTHR | 327F | Low | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific. | 5 days | |
| WTHR | 327F | Low | Open chemical containers not in use in fume hood B. | Immed | |
| WTHR | 327F | Low | Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls | Immed | |
| WTHR | 327F | Low | Container of chemical is unlabeled. (Unlabeled flask on bench top.) | Immed | |
| WTHR | 331C | Low | Heat resistant pad (ceramic tile or metal stove and counter top pad) needed under coffee maker(s) and/or hot water heater(s).* | 60 days | 12/12/02 |
| WTHR | 322 | Low | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific. | 5 days | |
| WTHR | 322 | Low | Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls | Immed | |
| WTHR | 333B | Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls | Immed | ||
| WTHR | 333B | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific. | 5 days | ||
| WTHR | 333B | Display written SOP for accessing extremely high containers of liquids, or move containers of liquids stored above eye level to eye level or below. | 30 days | ||
| WTHR | 333B | Chemical hood, glovebox, and/or laminar flow filtered air unit unsafe due to excessive accumulation of containers and other materials. Ether stored next to an ignition source in fume hood . | 10 days | ||
| WTHR | 333C | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. | 5 days | ||
| WTHR | 333C | Chemical hood, glovebox, and/or laminar flow filtered air unit unsafe due to excessive accumulation of containers and other materials. Ether stored next to an ignition source in fume hood C. | 10 days | ||
| WTHR | 352 hallway | Corridor obstructed. Hallway used as a storage area | 10 days | ||
| WTHR | 357 | Display written SOP for accessing extremely high containers of liquids, or move containers of liquids stored above eye level to eye level or below. | 30 days | ||
| WTHR | 357 | Chemical hood, glovebox, and/or laminar flow filtered air unit unsafe due to excessive accumulation of containers and other materials. Ethyl ether anhydrous & ethyl alcohol stored next to heater inside fume hood B. | 10 days | ||
| WTHR | 357 | Open chemical containers not in use. (In fume hood B) | Immed | ||
| WTHR | 357 | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific and clean copy. | 5 days | ||
| WTHR | 357 | BCHM div | Clean broken glass or plastic sharps should be collected in a cardboard box or other strong, secure disposable container. Needles, razor blades, scalpels, etc., must be disposed of in puncture proof plastic containers. Sharps container was over filled. | Immed | |
| WTHR | 365 | NMR | Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls. (Abbreviations on containers.) | Immed | |
| WTHR | 373 | Lipton | Failure to designate areas for select carcinogens, reproductive hazards, and/or extremely toxic chemicals. Use supplied "CAUTION" stickers as needed in these areas. See http://www.purdue.edu/rem/home/booklets/crdalist.doc (Chloroform) | 10 days | 12/16/2002 |
| WTHR | 373 | Lipton | Chemical hood, glovebox, and/or laminar flow filtered air unit unsafe due to excessive accumulation of containers and other materials. Ether stored next to an ignition source in fume hoods A & B. | 10 days | 12/18/2002 |
| WTHR | 373 | Lipton | Unsecured or inadequately secured compressed gas cylinder(s). Uncapped cylinders without regulators are unacceptable. Cylinders with regulators should be secured individually. (Hydrogen cylinder without a cap or regulator.) | Immed | 12/16/2002 |
| WTHR | 373 | Lipton | Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls | Immed | 12/16/2002 |
| WTHR | 373 | Lipton | Display written SOP for accessing extremely high containers of liquids, or move containers of liquids stored above eye level to eye level or below. | 30 days | 12/16/2002 |
| WTHR | 373A | Lipton | Container of chemical is unlabeled. Unlabeled flask in fume hood. | Immed | 12/16/2002 |
| WTHR | 373A | Lipton | Possible potable water supply cross connection exists. Hose(s) from spigot placed inside drain and/or touch bottom of basin. Cut hose(s) to a length no more than halfway below flood plane. | 30 days | 12/16/2002 |
| WTHR | 373A | Lipton | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. | 5 days | 12/16/2002 |
| WTHR | 373A | Lipton | Combustible materials not stored safely. Excessive empty box storage. Remove boxes or break boxes down for storage. | Immed | 12/16/2002 |
| WTHR | 377 | Lipton | Power cord damaged or broken. Electrical tape is unacceptable. Frayed cord on flash evaporator in fume hood C. | 10 days | 12/16/2002 |
| WTHR | 377 | Lipton | Display written SOP for accessing extremely high containers of liquids, or move containers of liquids stored above eye level to eye level or below. | 30 days | 12/16/2002 |
| WTHR | 377 | Lipton | Sign needed to indicate hazard. Signage present on fume hood but was obstructed by lab coats. | 30 days | 12/16/2002 |
| WTHR | 377 | Lipton | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific. | 5 days | 12/16/2002 |
| WTHR | 377 | Lipton | Open chemical containers not in use in fume hood C. | Immed | 12/16/2002 |
| WTHR | 377A | Lipton | Chemical hood, glovebox, and/or laminar flow filtered air unit unsafe due to excessive accumulation of containers and other materials. Acetone stored next to an ignition source in fume hood. | 10 days | 12/16/2002 |
| WTHR | 415 | Flammable storage cabinet closure and latch mechanisms defective, unreliable, needs repair work. | 60 days | ||
| WTHR | 419 | Lipton | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. | 5 days | 12/16/2002 |
| WTHR | 419 | Lipton | Glass vacuum lines, pressure lines, or Dewar flasks not taped or caged. | 10 days | 12/16/2002 |
| WTHR | 419 | Lipton | Excessive accumulation of chemicals given the available space and/or apparent storage and care conditions. Stacked storage in bottom of flam cab is NOT acceptable. | 30 days | 12/16/2002 |
| WTHR | 419 | Lipton | Display written SOP for accessing extremely high containers of liquids, or move containers of liquids stored above eye level to eye level or below. | 30 days | 12/16/2002 |
| WTHR | 419A | Lipton | Alcoholic beverages are stored in this area or there is an excessive accumulation of empty alcohol containers. And "wine tasting schedule" near by which did not identify any off-campus location. | Immed | 12/16/2002 |
| WTHR | 423A | Thompson | Hood must be capable of closing completely. Re-route lyophilizer cord. | Immed | |
| WTHR | 423A | Thompson | Excessive use of floor space for containers of chemicals, equipment, and/or other items restrict aisleways. Too easy to trip on glass disposal box near walk-in hood. | Immed | |
| WTHR | 423A | Thompson | Hazard assessment certification is posted but is not appropriate to this work area. (Welding!?) Assessment needs to be specific | 10 days | |
| WTHR | 430 | Morrison | First aid kits (which are not required in this area) must be posted with inventory, must contain exactly what the inventory claims, and may not contain any damaged items or items past their expiration date. (First aid kit at entry to 440.) | 30 days | 12/18/2002 |
| WTHR | 430B | Morrison | Hoods function properly when work is set up 6" from face opening. Add a tape or paint line to any/all hoods where necessary to remind users to keep the work 6" back. [there are no hoods in 430B, inspector/report error?] | 60 days | 10/28/2002 |
| WTHR | 430B | Morrison | No employee shall be allowed to consume food or beverages in …. any area exposed to a toxic material. The entire room is a "designated area" as per carcinogens…. Sticker on door. No eating or drinking of any kind may be done if entire room is designated | Immed | 12/13/2002 |
| WTHR | 430B | Morrison | Hazard assessment certification is posted but is not appropriate to this work area. Remove WELDING and any other sections not necessary. Ensure that rules are clear, appropriate and enforced. | 10 days | 12/19/2002 |
| WTHR | 430D | Morrison | Hazard assessment certification is posted but is not appropriate to this work area. Remove WELDING and any other sections not necessary. Ensure that rules are clear, appropriate and enforced. | 10 days | 12/19/2002 |
| WTHR | 440 | Morrison | Unsecured or inadequately secured compressed gas cylinder(s). Uncapped cylinders without regulators are unacceptable. Cylinders with regulators should be secured individually. (Corrected immedly.) | Immed | 10/28/2002 |
| WTHR | 440 | Wei | Excess flammables stored outside of approved cabinet and/or safety cans. | 60 days | |
| WTHR | 440 | Wei | Chemical hood, glovebox, and/or laminar flow filtered air unit unsafe due to excessive accumulation of containers and other materials. Institute a 6-inch (tape) line and rule to avoid chemicals used too near sash. | 10 days | |
| WTHR | 440A | Morrison | Power cord damaged or broken. Electrical tape is unacceptable. (Frayed cloth-braided cord plugged into Variac.) | 10 days | 12/19/2002 |
| WTHR | 440A | Morrison | Chemical spills on the floor and/or working surfaces must be cleaned. (Pump oil on benchtop.) | 10 days | 12/19/2002 |
| WTHR | 440B | Morrison | Fire doors must not be left propped open when room is unattended. Door between 440 and 440B was propped open during early morning, no one at work yet. | Immed | 12/18/2002 |
| WTHR | 440B | Morrison | Excess flammables stored outside of approved cabinet and/or safety cans. | 60 days | 12/18/2002 |
| WTHR | 440B | Morrison | Open chemical containers not in use, in hood C. | Immed | 12/13/2002 |
| WTHR | 440B | Morrison | Containers of chemicals are inadequately labeled (hood C, e.g.). If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls | Immed | 12/13/2002 |
| WTHR | 440B | Morrison | Hazard assessment certification posted but is not appropriate to this work area. (Welding!?) . Assessment needs to be specific. | 10 days | 12/13/2002 |
| WTHR | 441 | Grutzner | Post entry into 441B as "NOT AN EXIT." | 30 days | |
| WTHR | 441 | Grutzner | No employee shall be allowed to consume food or beverages in …. any area exposed to a toxic material. There is evidence of food and drink being consumed much too close to where chemicals are used/stored (on desks). | Immed | |
| WTHR | 441 | Grutzner | Restore housekeeping. Severe problem with many items sticking out from benchtops into walking areas. | 10 days | |
| WTHR | 441 | Grutzner | Fire doors must not be left propped open when room is unattended. Both doors between 441 and 463 are to be kept closed. Never TIE open any fire door. | Immed | |
| WTHR | 441 | Grutzner | Excessive use of floor space for containers of chemicals, equipment, and/or other items restrict aisleways. | Immed | |
| WTHR | 442 | Fuchs | Excess flammables stored outside of approved cabinet and/or safety cans. | 60 days | 12/19/2002 |
| WTHR | 442 | Fuchs | Hoods function properly when work is set up 6" from face opening. Add a tape or paint line to any/all hoods where necessary to remind users to keep the work 6" back from sash. | 60 days | 1/7/2003 |
| WTHR | 442 | Fuchs | Excessive use of floor space for containers of chemicals, equipment, and/or other items restrict aisleways. | Immed | 12/19/02 |
| WTHR | 442 | Fuchs | Hazard assessment certification is posted but is not appropriate to this work area. Remove WELDING and any other sections not necessary. Ensure that rules are clear, appropriate and enforced. | 10 days | 1/24/03 |
| WTHR | 442 | Fuchs | Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls | Immed | 12/19/2002 |
| WTHR | 450A | Fuchs | Excessive use of floor space for containers of chemicals, equipment, and/or other items restrict aisleways. | Immed | 12/19/2002 |
| WTHR | 450A | Fuchs | Hoods function properly when work is set up 6" from face opening. Add a tape or paint line to any/all hoods where necessary to remind users to keep the work 6" back from sash. | 60 days | 1/7/03 |
| WTHR | 450D | Fuchs | Hoods function properly when work is set up 6" from face opening. Add a tape or paint line to any/all hoods where necessary to remind users to keep the work 6" back from sash. | 60 days | 1/7/03 |
| WTHR | 450D | Fuchs | Hazard assessment certification is posted but is not appropriate to this work area. Remove WELDING and any other sections not necessary. Ensure that rules are clear, appropriate and enforced. | 10 days | 1/24/03 |
| WTHR | 450E | Fuchs | Excessive use of floor space for containers of chemicals, equipment, and/or other items restrict aisleways. | Immed |
12/19/02 |
| WTHR | 450E | Fuchs | Hoods function properly when work is set up 6" from face opening. Add a tape or paint line to any/all hoods where necessary to remind users to keep the work 6" back from sash. | 60 days | 1/7/03 |
| WTHR | 454 | Negishi | Workplace not reasonably free of recognized hazards. Liquid in refrigerator without secure closure, parafilm is NOT allowed as only closure for liquids in refrigerators. | 10 days | 12/23/2002 |
| WTHR | 454 | Negishi | Hoods function properly when work is set up 6" from face opening. Add a tape or paint line to any/all hoods where necessary to remind users to keep the work 6" back from sash. | 60 days | 1/10/2003 |
| WTHR | 454 | Negishi | Excessive use of floor space for containers of chemicals, equipment, and/or other items restrict aisleways. (Although in general the lab and benches look very tidy.) | Immed | 12/14/2002 |
| WTHR | 463 | Grutzner | Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. | 5 days | |
| WTHR | 463 | Grutzner | Excessive use of floor space for containers of chemicals, equipment, and/or other items that restrict aisleways. | Immed | |
| WTHR | 463 | Grutzner | Excess flammables stored outside of approved cabinet and/or safety cans. | 60 days | |
| WTHR | 465 | Negishi | Excessive use of floor space for containers of chemicals, equipment, and/or other items restrict aisleways. | Immed | 12/14/2002 |
| WTHR | 465 | Negishi | Restore housekeeping. Severe problem with many items sticking out from benchtops into walking areas. | 10 days | 12/20/2002 |
| WTHR | 465 | Negishi | Hoods function properly when work is set up 6" from face opening. Add a tape or paint line to any/all hoods where necessary to remind users to keep the work 6" back from sash. | 60 days | 1/10/2003 |
| WTHR | 465 | Negishi | Hazard assessment certification is posted but refers to a non-existent page 2. Personal protective equipment must be specific. | 30 days | 1/10/2003 |
| WTHR | 465 | Negishi | Open chemical containers not in use. | Immed | 12/14/2002 |