Get RTF file here  and follow the instructions.  Depending on your configuration, a right click on that link might open it or might produce a "do you want to save" message.  Save it to a local or network drive and keep track of it.  If your file extensions are showing and you don't change the name, it is called BRWN_WTHR_2002.rtf  Any recent word processor version on a PC or Mac should open it.

bldg

room

responsible

Item of concern

abatement time

date corrected
BRWN B125A Wenthold Workplace not reasonably free of recognized hazards. Compressed gas cylinder containing 5% fluorine:95% helium not in vented gas cabinet. 60 days  
BRWN B137 Ben-Amotz Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Specific PPE requirements needed instead of generic. 5 days 1/13/2003
BRWN B165A Francisco(?) Unsecured or inadequately secured compressed gas cylinder(s). Uncapped cylinders without regulators are unacceptable. Cylinders with regulators should be secured individually. Small cylinder not secured. Immed  
BRWN B178 McLuckey Flexible cords and/or cables not protected from accidental damage. Power cord across walkway needs to be removed. Cord tunnel not acceptable. Immed 1/17/2003
BRWN Basement
Corridor
Dirksen Corridor is obstructed by 42 pallets of office partitions and assembly parts. Items need to be removed from the corridor. 60 days 1/16/2003
BRWN 1142 Swihart Workplace not reasonably free of recognized hazards: duplex outlet in kitchenette near sink is without GFCI protection. Submit OSHA 113 to REM. REM will issue 18-A. P59 is not required. 60 days 113 submitted 1/24/2003
BRWN 1152B   Restore aisleway. Grocery sacks of material, carpeting roll, cloth, and board obstructs the aisleway. 10 days  
BRWN 2105   Restore aisleway. 10 days  
BRWN 2105   Heat resistant pad (ceramic tile or metal stove and counter top pad) needed under coffee maker(s) and/or hot water heater(s). (Needed for large coffee maker) 60 days  
BRWN 2105   Electrical cord tunnels across aisleways are not acceptable. 30 days  
BRWN 2107B Oakley Heat resistant pad (ceramic tile or metal stove and counter top pad) needed under coffee maker(s) and/or hot water heater(s). 60 days 12/12/02
BRWN 2184 Keyes Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC 10 days 12/12/2002
BRWN 2184 Keyes Display written SOP for accessing extremely high containers of liquids, or move containers of liquids stored above eye level to eye level or below. 30 days 12/13/2002
BRWN 2184 Keyes Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. 5 days 12/12//2003
BRWN 3124 Hrycyna Eyewash and/or shower is obstructed. Immed 1/9/2003
BRWN 3124 Hrycyna Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls Immed 1/9/2003
BRWN 3124 Hrycyna Unsecured or inadequately secured compressed gas cylinder(s). Uncapped cylinders without regulators are unacceptabl. (3 cylinders without regulators ganged together.) Immed 1/9/2003
BRWN 3124 Hrycyna Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific. 5 days 1/9/2003
BRWN 3124 Hrycyna Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC 10 days 1/9/2003
BRWN 3124A Laskowski Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC 10 days 12/20/2002
BRWN 3124A Laskowski Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific. 5 days 12/20/2002
BRWN 3124A Laskowski Failure to designate areas for select carcinogens, reproductive hazards, and/or extremely toxic chemicals. Use supplied "CAUTION" stickers as needed in these areas. See http://www.purdue.edu/rem/home/booklets/crdalist.doc (Chloroform) 10 days 12/20/2002
BRWN 3124B Hrycyna Eyewash and/or shower is obstructed. Immed 1/9/2003
BRWN 3124B Hrycyna Excessive fire fuel loading (paper and/or other combustibles). Excessive accumulation of styrofoam containers. Immed 1/9/2003
BRWN 3124B Hrycyna Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls Immed 1/9/2003
BRWN 3124B Hrycyna Display written SOP for accessing extremely high containers of liquids, or move containers of liquids stored above eye level to eye level or below. 30 days 1/9/2003
BRWN 3124B Hrycyna Open chemical containers not in use. (5 gallon carboy uncapped.) Immed 1/9/2003
BRWN 3124B Hrycyna Container of chemical is unlabeled. (5 gallon carboy unlabeled.) Immed 1/9/2003
BRWN 3124B Hrycyna Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific. 5 days 1/9/2003
BRWN 3124B Hrycyna Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC 10 days 1/9/2003
BRWN 3125A Chmielewski Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC 10 days 12/17/2002
BRWN 3125A Chmielewski Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. 5 days 12/16/2002
BRWN 3125B Chmielewski Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls Immed 12/17/2002
BRWN 3134 Laskowski Container of chemical is unlabeled. Immed 12/20/2002
BRWN 3134 Laskowski Combustible materials not stored safely. (Excessive saw dust accumulation in fume hood B.) Immed 12/20/2002
BRWN 3134 Laskowski Chemical hood, glovebox, and/or laminar flow filtered air unit unsafe due to excessive accumulation of containers and other materials. Flammables stored next to an ignition source in fume hood A. 10 days 12/20/2002
BRWN 3134A Laskowski Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. 5 days 12/20/2002
BRWN 3134A Laskowski Food stored with laboratory equipment and/or chemicals in walk-in. Immed 12/13/2002
BRWN 3134B Laskowski Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific. 5 days 12/20/2002
BRWN 3134C Laskowski Display written SOP for accessing extremely high containers of liquids, or move containers of liquids stored above eye level to eye level or below. 30 days 12/20/2002
BRWN 3134C Laskowski Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific. 5 days 12/20/2002
BRWN 3134C Laskowski Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC 10 days 12/20/2002
BRWN 3134E   Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific. 5 days  
BRWN 3135   Container of chemical is unlabeled. (Carboy unlabeled.) Immed  
BRWN 3135A   Eyewash and/or shower is obstructed. Immed  
BRWN 3135E   Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. 5 days  
BRWN 3144 Laskowski Container of chemical is unlabeled. Immed 12/20/2002
BRWN 3144 Laskowski Restore aisleway. 10 days 12/20/2002
BRWN 3144 Laskowski Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific. 5 days 12/20/2002
BRWN 3151A corridor   Corridor obstructed with bookcases and filing cabinet. 10 days  
BRWN 3151E   Eyewash and/or shower is obstructed. Eyewash is obstructed by curtain on safety shower. Immed  
BRWN 3151E   Electrical cord tunnels across aisleways are not acceptable. 30 days  
BRWN 3151E   Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC 10 days  
BRWN 3151F   Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC 10 days  
BRWN 3151F   Chemical hood, glovebox, and/or laminar flow filtered air unit unsafe due to excessive accumulation of containers and other materials. Flammables stored next to an ignition source in both fume hoods. 10 days  
BRWN 3164   Eyewash and/or shower is obstructed. Immed  
BRWN 3164   Container of chemical is unlabeled. Immed  
BRWN 3164   Display written SOP for accessing extremely high containers of liquids, or move containers of liquids stored above eye level to eye level or below. 30 days  
BRWN 3164   Failure to designate flam. storage cabinets for select carcinogens, reproductive hazards, and/or extremely toxic chemicals. Use supplied "CAUTION" stickers as needed in these areas. See http://www.purdue.edu/rem/home/booklets/crdalist.doc 10 days  
BRWN 3164   Unsecured or inadequately secured compressed gas cylinder(s). Uncapped cylinders without regulators are unacceptable. Cylinders with regulators should be secured individually. (Improperly secured helium cylinder.) Immed  
BRWN 3164   Electrical equipment shall be free from recognized hazards that are likely to cause death or serious physical harm to employees. (Hot plate is missing the base plate.) Immed  
BRWN 3170   Unsecured or inadequately secured compressed gas cylinder(s). Uncapped cylinders without regulators are unacceptable. Cylinders with regulators should be secured individually. (4 unsecured cylinders.) Immed  
BRWN 3171   Container of chemical is unlabeled. Immed  
BRWN 3171   Unsecured or inadequately secured compressed gas cylinder(s). Uncapped cylinders without regulators are unacceptable. Cylinders with regulators should be secured individually. Immed  
BRWN 3174   Clean broken glass or plastic sharps should be collected in a cardboard box or other strong, secure disposable container. Needles, razor blades, scalpels, etc., must be disposed of in puncture proof plastic containers. Sharps container was over filled. Immed  
BRWN 3174   Restore aisleway. (CORRECTED by REM.) 10 days  
BRWN 3174   Display written SOP for accessing extremely high containers of liquids, or move containers of liquids stored above eye level to eye level or below. 30 days  
BRWN 3174   Chemical containers have labels that are deteriorating and/or unreadable in fume hood A. Immed  
BRWN 3175 Keyes Power cord damaged or broken. Electrical tape is unacceptable. (End of electrical cord on Thelco brand lab equipment is broken.) 10 days 12/13/2002
BRWN 3175 Keyes Container of chemical is unlabeled. Immed 12/13/2002
BRWN 3182   Display written SOP for accessing extremely high containers of liquids, or move containers of liquids stored above eye level to eye level or below. 30 days 12/12/2002
BRWN 5125 Ramachandran Open chemical containers not in use. Immed 12/12/2002
BRWN 5125 Ramachandran Excess flammables stored outside of approved cabinet and/or safety cans. There are flam. cabs used for non flam. storage and excess flam. liquids outside of appropriate storage. Only 10 gal total Class I plus Class II liquids outside ot flam. storage. 60 days 12/12/2002
BRWN 5125 Ramachandran Flammable storage cabinet not closed properly. Must be kept closed to be effective. Corrected at inspection. Immed 10/28/2002
BRWN 5125 Ramachandran Excessive use of floor space for containers of chemicals, equipment, and/or other items restrict aisleways. Remove ALL chemicals and waste from walking and working aisles, also equipment, also cans of used syringes and needles, also safety cans. Immed 12/15/2002
BRWN 5125 Ramachandran Excessive accumulation or abandoned containers of chemical waste. 10 days 12/13/2002
BRWN 5125 Ramachandran Write date on waste containers at start of accumulation. Waste is very poorly managed in this room and this step is required that more attention be paid to it. Immed 12/16/2002
BRWN 5125 Ramachandran Employee observed working with liquids without splash goggles, required by sign on hallway side of 5125 sign. Immed 12/12/2002
BRWN 5125 Ramachandran Hazard assessment certification is posted but is not appropriate to this work area. Remove WELDING and any other sections not necessary. Ensure that rules are clear, appropriate and enforced. 10 days 1/16/2003
BRWN 5125 Ramachandran No employee shall be allowed to consume food or beverages … in any area exposed to a toxic material. This room shall be posted completely off-limits for all food and beverage. Immed 12/16/2002
BRWN 5125 Ramachandran Hoods function properly when work is set up 6" from face opening. Add a tape or paint line to any/all hoods where necessary to remind users to keep the work 6" back. 60 days 12/12/2002
BRWN 5129 Ramachandran Unsecured compressed gas cylinder(s), also other cylinders ganged behind a single support chain. Cylinders should be secured individually while cap is off. Immed  
BRWN 5131 Ramachandran Failure to use required personal protective equipment. Main entance sign requires wearing of goggles inside at all times. Remove sign or enforce. Immed 12/12/2002
BRWN 5134 Rothwell Excess flammables stored outside of approved cabinet and/or safety cans. A flam cab exists which contains many NON flammables. Take out nonflammables, put in flammables. 60 days  
BRWN 5134 Rothwell Hazard assessment certification is posted but is not appropriate to this work area. Remove WELDING and any other sections not necessary. Ensure that rules are clear, appropriate and enforced. 10 days  
BRWN 5134A Rothwell Open chemical waste containers not in use. Funnel top cover was in up position. Corrected at inspection. Immed  
BRWN 5135 Ramachandran Excessive use of floor space for containers of chemicals, equipment, and/or other items restrict aisleways. Remove ALL chemicals and waste from walking and working aisles, also equipment, also cans of used syringes and needles, also safety cans. Immed 12/13/2002
BRWN 5135 Ramachandran Write date on waste containers at start of accumulation. Waste is very poorly managed in this room and this step is required that more attention be paid to it. Immed 12/14/2002
BRWN 5135 Ramachandran Hoods function properly when work is set up 6" from face opening. Add a tape or paint line to any/all hoods where necessary to remind users to keep the work 6" back from face of hood. 60 days 12/12/2002
BRWN 5135 Ramachandran No employee shall be allowed to consume food or beverages … in any area exposed to a toxic material. This room shall be posted completely off-limits for all food and beverage. Immed 12/12/2002
BRWN 5135 Ramachandran Excessive accumulation or abandoned containers of chemical waste. Nearly 50 gallons of waste in over 15 containers, mostly on the floor. Waste MUST be managed more effectively. 10 days 12/13/2002
BRWN 5135 Ramachandran Excess flammables stored outside of approved cabinet and/or safety cans. Flam cab in this room is extremely under-utilized, could take more flammables if shelves were replaced. 60 days 12/12/2002
BRWN 5135 Ramachandran Hazard assessment certification is posted but is not appropriate to this work area. Remove WELDING and any other sections not necessary. Ensure that rules are clear, appropriate and enforced. 10 days 1/16/200
3BRWN 5135 Ramachandran Failure to use personal protective equipment required by hazard assessment certification. Employee observed pouring liquids without splash goggles. Immed 12/12/2002
BRWN 5150 Fenniri Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls Immed  
BRWN 5150 Fenniri Flammable storage cabinet not closed properly. Must be kept closed to be effective. Corrected at inspection. variable  
BRWN 5150 Fenniri Hoods function properly when work is set up 6" from face opening. Add a tape or paint line to any/all hoods where necessary to remind users to keep the work 6" back. 60 days  
BRWN 5150 Fenniri Excess flammables stored outside of approved cabinet and/or safety cans. 60 days  
BRWN 5150   Excessive use of floor space for containers of chemicals, equipment, and/or other items restrict aisleways. Immed  
WTHR 011, Stairwell nearby   Fire doors must not be left propped open to this stairwell. Immed  
WTHR 012 Dirksen Restore aisleways. Aisleway blocked on the south side and the east side of the room. 10 days 1/16/2003
WTHR 031 Cooks Workplace not reasonably free of recognized hazards: duplex outlet in kitchenette without GFCI protection. Submit OSHA 113 to REM. REM will issue 18-A. P59 is not required. 60 days 113 submitted 1/24/2003
WTHR 033C Cooks Fire doors must not be left propped open when room is unattended. Twin firedoors chocked open to 033C. Immed 12/16/2002
WTHR 033C Cooks Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC 10 days 12/18/2002
WTHR 033D Cooks Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC 10 days 12/18/2002
WTHR 040   No backflow preventers or flashback arresters installed on oxygen and acetylene tank of welding outfit. One of each needed at each tank outlet. 60 days  
WTHR 055A   Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC 10 days  
WTHR 060   Container of chemical is unlabeled. Five gallon carboy on west shelf containing a clear liquid is without contents and hazard labels. Immed  
WTHR 062 Cooks Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls Immed 12/18/2002
WTHR 062 Cooks Open chemical containers not in use. Funnel in 5 gallon methylene chloride open container. Immed 12/17/2002
WTHR 062 Cooks Flexible cords and/or cables not protected from accidental damage. Computer cable across walkway at NE door to room needs to be removed. Immed 1/03/2003
WTHR 070   Container of chemical is unlabeled. Spray bottle with blue liquid is without label of contents and hazards. Immed  
WTHR 075   Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC 10 days  
WTHR 077 Cooks Unsecured or inadequately secured compressed gas cylinder(s). Uncapped cylinders without regulators are unacceptable. Cylinders with regulators should be secured individually. Immed 12/18/2002
WTHR 081   Container of chemical is unlabeled. Squeeze bottle with clear liquid is without contents and hazard labels. Immed  
WTHR 081   Food stored with laboratory equipment and/or chemicals. Food stored in chemical refrigerator with chemicals. Immed  
WTHR 083, in corridor Dirksen Corridor obstructed by 13 desks and 7 benches between elevator and room 083. 10 days 1/17/2003
WTHR 101   Flexible cords and/or cables not protected from accidental damage. Tubing across walkway needs to be removed. Immed  
WTHR 105   Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. 5 days  
WTHR 110B Lipschutz Failure to designate areas for select carcinogens, reproductive hazards, and/or extremely toxic chemicals. Use supplied "CAUTION" stickers as needed in these areas. See http://www.purdue.edu/rem/home/booklets/crdalist.doc ( Formaldehyde) 10 days  
WTHR 110B Lipschutz Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Specific not generic assessment needed. 5 days  
WTHR 117A Nash Heat resistant pad (ceramic tile or metal stove and counter top pad) needed under coffee maker.* 60 days 12/16/2002
WTHR 120A   Interference to sprinkler discharge because of building contents. Minimum vertical clearance between sprinklers and material below shall be 18": cardboard boxes on the top row. Immed  
WTHR 127   Combustible materials not stored safely. More than 100 empty cardboard boxes stored in north central part of room need to be collapsed for storage. Immed  
WTHR 141 (140A*) Dirksen Grinder's work rest is greater than 0.125" from abrasive wheel. One wheel is missing tool rest on 1/2 H.P. Sears bench grinder. 10 days 1/17/2003
WTHR 144 Rothwell/Wood Unsecured or inadequately secured compressed gas cylinder(s). Uncapped cylinders without regulators are unacceptable. Cylinders with regulators should be secured individually. Immed 12/13/2002
WTHR 144 Rothwell/Wood Container of chemical is unlabeled. Squeeze bottle on sink is without contents and hazard labels. Immed 12/13/2002
WTHR 165   Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Specific not generic PPE assessment needed. 5 days 12/2/2002

 

WTHR 168   Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Specific not generic PPE assessment needed. 5 days 12/2/2002

 

WTHR 173B Paul Bower Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Specific not generic PPE assessment needed. 5 days 12/2/2002
WTHR 173B Paul Bower Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC 10 days 12/20/2002
WTHR 175   Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC 10 days 1/9/03
WTHR 175   Unsecured or inadequately secured compressed gas cylinder(s). Uncapped cylinders without regulators are unacceptable. Cylinders with regulators should be secured individually. At west end of room, 2 cylinders inadequately secured. Immed 12/2/2002
WTHR 175   Failure to designate areas for select carcinogens, reproductive hazards, and/or extremely toxic chemicals. Use supplied "CAUTION" stickers as needed in these areas. See http://www.purdue.edu/rem/home/booklets/crdalist.doc (Toluene) 10 days 12/20/2002
WTHR 175   Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. 5 days 12/20/2002
WTHR 185A   Restore aisleway. East and west door aisles obstructed. 10 days  
WTHR 206   Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific. 5 days  
WTHR 206   Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC 10 days  
WTHR 206   Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls Immed  
WTHR 206   Container of chemical is unlabeled. Unlabeled bottles in fume hoods A & B. Immed  
WTHR 208   Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls Immed  
WTHR 208   Failure to designate areas for select carcinogens, reproductive hazards, and/or extremely toxic chemicals. Use supplied "CAUTION" stickers as needed in these areas. See http://www.purdue.edu/rem/home/booklets/crdalist.doc 10 days  
WTHR 208   Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific. 5 days  
WTHR 208   Emergency telephone numbers (24 hours) must be posted on outside of lab door(s). See http://www.purdue.edu/rem/home/forms/DOORPOST.DOC 10 days  
WTHR 217   Heat resistant pad (ceramic tile or metal stove and counter top pad) needed under coffee maker(s) and/or hot water heater(s).* 60 days  
WTHR 221   Heat resistant pad (ceramic tile or metal stove and counter top pad) needed under coffee maker(s) and/or hot water heater(s).* 60 days  
WTHR 221B   Heat resistant pad (ceramic tile or metal stove and counter top pad) needed under coffee maker(s) and/or hot water heater(s).* 60 days  
WTHR 225   Electrical cord tunnels across aisleways are not acceptable. 30 days 1/31/2003
WTHR 227A   Heat resistant pad (ceramic tile or metal stove and counter top pad) needed under coffee maker(s) and/or hot water heater(s).* 60 days 1/31/2003
WTHR 236   Required exit(s) obstructed. Exit to hallway is obstructed. Immed  
WTHR 242   Container of chemical is unlabeled. Immed  
WTHR 242   Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls Immed  
WTHR 242   Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. 5 days  
WTHR 248   Container of chemical is unlabeled in fume hood A. Immed  
WTHR 248   Open chemical containers not in use in fume hood A. Immed  
WTHR 250A   Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. 5 days  
WTHR 250A   Container of chemical is unlabeled. Immed  
WTHR 250B   Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls Immed  
WTHR 265   Heat resistant pad (ceramic tile or metal stove and counter top pad) needed under coffee maker(s) and/or hot water heater(s). (3 needed)* 60 days  
WTHR 316   Container of chemical is unlabeled. (Unlabeled container on benchtop.) Immed  
WTHR 317   Container of chemical is unlabeled. (Unlabeled container on benchtop.) Immed  
WTHR 327 Low No employee shall be allowed to consume food or beverages in a toilet room nor in any area exposed to a toxic material. Evidence of drinking in Biohazard designated area. Immed 1/21/2002
WTHR 327 Low Clean broken glass or plastic sharps should be collected in a cardboard box or other strong, secure disposable container. Needles, razor blades, scalpels, etc., must be disposed of in puncture proof plastic containers. Sharps container was over filled. Immed 1/17/2003
WTHR 327 Low Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific and clean copy. 5 days 1/21/2003
WTHR 327 Low Unsecured or inadequately secured compressed gas cylinder(s). Uncapped cylinders without regulators are unacceptable. Cylinders with regulators should be secured individually. (Carbon dioxide cylinder was without a cap or regulator.)  Corrected at inspection. Immed 10/28/2002
WTHR 327 Low Container of chemical is unlabeled. (In Biosafety Cabinet.) Immed 1/17/2003
WTHR 327 Low Open chemical containers not in use. (In Biosafety Cabinet.) Immed 1/17/2003
WTHR 327 Low Display written SOP for accessing extremely high containers of liquids, or move containers of liquids stored above eye level to eye level or below. 30 days 1/17/2003
WTHR 327 Low Stored chemicals are not segregated by hazards. (Hood A - perchloric acid stored with acetic acid.) Immed  
WTHR 327 Low Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls Immed  
WTHR 327B Low Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. 5 days 1/21/2003
WTHR 327C Low Heat resistant pad (ceramic tile or metal stove and counter top pad) needed under coffee maker(s) and/or hot water heater(s).* 60 days Kevin fixed 331C, not this one
WTHR 327E Low Open chemical containers not in use. Immed 1/17/2003
WTHR 327F Low Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific. 5 days  
WTHR 327F Low Open chemical containers not in use in fume hood B. Immed  
WTHR 327F Low Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls Immed  
WTHR 327F Low Container of chemical is unlabeled. (Unlabeled flask on bench top.) Immed  
WTHR 331C Low Heat resistant pad (ceramic tile or metal stove and counter top pad) needed under coffee maker(s) and/or hot water heater(s).* 60 days 12/12/02
WTHR 322 Low Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific. 5 days  
WTHR 322 Low Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls Immed  
WTHR 333B   Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls Immed  
WTHR 333B   Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific. 5 days  
WTHR 333B   Display written SOP for accessing extremely high containers of liquids, or move containers of liquids stored above eye level to eye level or below. 30 days  
WTHR 333B   Chemical hood, glovebox, and/or laminar flow filtered air unit unsafe due to excessive accumulation of containers and other materials. Ether stored next to an ignition source in fume hood . 10 days  
WTHR 333C   Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. 5 days  
WTHR 333C   Chemical hood, glovebox, and/or laminar flow filtered air unit unsafe due to excessive accumulation of containers and other materials. Ether stored next to an ignition source in fume hood C. 10 days  
WTHR 352 hallway   Corridor obstructed. Hallway used as a storage area 10 days  
WTHR 357   Display written SOP for accessing extremely high containers of liquids, or move containers of liquids stored above eye level to eye level or below. 30 days  
WTHR 357   Chemical hood, glovebox, and/or laminar flow filtered air unit unsafe due to excessive accumulation of containers and other materials. Ethyl ether anhydrous & ethyl alcohol stored next to heater inside fume hood B. 10 days  
WTHR 357   Open chemical containers not in use. (In fume hood B) Immed  
WTHR 357   Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific and clean copy. 5 days  
WTHR 357 BCHM div Clean broken glass or plastic sharps should be collected in a cardboard box or other strong, secure disposable container. Needles, razor blades, scalpels, etc., must be disposed of in puncture proof plastic containers. Sharps container was over filled. Immed  
WTHR 365 NMR Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls. (Abbreviations on containers.) Immed  
WTHR 373 Lipton Failure to designate areas for select carcinogens, reproductive hazards, and/or extremely toxic chemicals. Use supplied "CAUTION" stickers as needed in these areas. See http://www.purdue.edu/rem/home/booklets/crdalist.doc (Chloroform) 10 days 12/16/2002
WTHR 373 Lipton Chemical hood, glovebox, and/or laminar flow filtered air unit unsafe due to excessive accumulation of containers and other materials. Ether stored next to an ignition source in fume hoods A & B. 10 days 12/18/2002
WTHR 373 Lipton Unsecured or inadequately secured compressed gas cylinder(s). Uncapped cylinders without regulators are unacceptable. Cylinders with regulators should be secured individually. (Hydrogen cylinder without a cap or regulator.) Immed 12/16/2002
WTHR 373 Lipton Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls Immed 12/16/2002
WTHR 373 Lipton Display written SOP for accessing extremely high containers of liquids, or move containers of liquids stored above eye level to eye level or below. 30 days 12/16/2002
WTHR 373A Lipton Container of chemical is unlabeled. Unlabeled flask in fume hood. Immed 12/16/2002
WTHR 373A Lipton Possible potable water supply cross connection exists. Hose(s) from spigot placed inside drain and/or touch bottom of basin. Cut hose(s) to a length no more than halfway below flood plane. 30 days 12/16/2002
WTHR 373A Lipton Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. 5 days 12/16/2002
WTHR 373A Lipton Combustible materials not stored safely. Excessive empty box storage. Remove boxes or break boxes down for storage. Immed 12/16/2002
WTHR 377 Lipton Power cord damaged or broken. Electrical tape is unacceptable. Frayed cord on flash evaporator in fume hood C. 10 days 12/16/2002
WTHR 377 Lipton Display written SOP for accessing extremely high containers of liquids, or move containers of liquids stored above eye level to eye level or below. 30 days 12/16/2002
WTHR 377 Lipton Sign needed to indicate hazard. Signage present on fume hood but was obstructed by lab coats. 30 days 12/16/2002
WTHR 377 Lipton Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. Needs to be specific. 5 days 12/16/2002
WTHR 377 Lipton Open chemical containers not in use in fume hood C. Immed 12/16/2002
WTHR 377A Lipton Chemical hood, glovebox, and/or laminar flow filtered air unit unsafe due to excessive accumulation of containers and other materials. Acetone stored next to an ignition source in fume hood. 10 days 12/16/2002
WTHR 415   Flammable storage cabinet closure and latch mechanisms defective, unreliable, needs repair work. 60 days  
WTHR 419 Lipton Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. 5 days 12/16/2002
WTHR 419 Lipton Glass vacuum lines, pressure lines, or Dewar flasks not taped or caged. 10 days 12/16/2002
WTHR 419 Lipton Excessive accumulation of chemicals given the available space and/or apparent storage and care conditions. Stacked storage in bottom of flam cab is NOT acceptable. 30 days 12/16/2002
WTHR 419 Lipton Display written SOP for accessing extremely high containers of liquids, or move containers of liquids stored above eye level to eye level or below. 30 days 12/16/2002
WTHR 419A Lipton Alcoholic beverages are stored in this area or there is an excessive accumulation of empty alcohol containers. And "wine tasting schedule" near by which did not identify any off-campus location. Immed 12/16/2002
WTHR 423A Thompson Hood must be capable of closing completely. Re-route lyophilizer cord. Immed  
WTHR 423A Thompson Excessive use of floor space for containers of chemicals, equipment, and/or other items restrict aisleways. Too easy to trip on glass disposal box near walk-in hood. Immed  
WTHR 423A Thompson Hazard assessment certification is posted but is not appropriate to this work area. (Welding!?) Assessment needs to be specific 10 days  
WTHR 430 Morrison First aid kits (which are not required in this area) must be posted with inventory, must contain exactly what the inventory claims, and may not contain any damaged items or items past their expiration date. (First aid kit at entry to 440.) 30 days 12/18/2002
WTHR 430B Morrison Hoods function properly when work is set up 6" from face opening. Add a tape or paint line to any/all hoods where necessary to remind users to keep the work 6" back.  [there are no hoods in 430B, inspector/report error?] 60 days 10/28/2002
WTHR 430B Morrison No employee shall be allowed to consume food or beverages in …. any area exposed to a toxic material. The entire room is a "designated area" as per carcinogens…. Sticker on door. No eating or drinking of any kind may be done if entire room is designated Immed 12/13/2002
WTHR 430B Morrison Hazard assessment certification is posted but is not appropriate to this work area. Remove WELDING and any other sections not necessary. Ensure that rules are clear, appropriate and enforced. 10 days 12/19/2002
WTHR 430D Morrison Hazard assessment certification is posted but is not appropriate to this work area. Remove WELDING and any other sections not necessary. Ensure that rules are clear, appropriate and enforced. 10 days 12/19/2002
WTHR 440 Morrison Unsecured or inadequately secured compressed gas cylinder(s). Uncapped cylinders without regulators are unacceptable. Cylinders with regulators should be secured individually. (Corrected immedly.) Immed 10/28/2002
WTHR 440 Wei Excess flammables stored outside of approved cabinet and/or safety cans. 60 days  
WTHR 440 Wei Chemical hood, glovebox, and/or laminar flow filtered air unit unsafe due to excessive accumulation of containers and other materials. Institute a 6-inch (tape) line and rule to avoid chemicals used too near sash. 10 days  
WTHR 440A Morrison Power cord damaged or broken. Electrical tape is unacceptable. (Frayed cloth-braided cord plugged into Variac.) 10 days 12/19/2002
WTHR 440A Morrison Chemical spills on the floor and/or working surfaces must be cleaned. (Pump oil on benchtop.) 10 days 12/19/2002
WTHR 440B Morrison Fire doors must not be left propped open when room is unattended. Door between 440 and 440B was propped open during early morning, no one at work yet. Immed 12/18/2002
WTHR 440B Morrison Excess flammables stored outside of approved cabinet and/or safety cans. 60 days 12/18/2002
WTHR 440B Morrison Open chemical containers not in use, in hood C. Immed 12/13/2002
WTHR 440B Morrison Containers of chemicals are inadequately labeled (hood C, e.g.). If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls Immed 12/13/2002
WTHR 440B Morrison Hazard assessment certification posted but is not appropriate to this work area. (Welding!?) . Assessment needs to be specific. 10 days 12/13/2002
WTHR 441 Grutzner Post entry into 441B as "NOT AN EXIT." 30 days  
WTHR 441 Grutzner No employee shall be allowed to consume food or beverages in …. any area exposed to a toxic material. There is evidence of food and drink being consumed much too close to where chemicals are used/stored (on desks). Immed  
WTHR 441 Grutzner Restore housekeeping. Severe problem with many items sticking out from benchtops into walking areas. 10 days  
WTHR 441 Grutzner Fire doors must not be left propped open when room is unattended. Both doors between 441 and 463 are to be kept closed. Never TIE open any fire door. Immed  
WTHR 441 Grutzner Excessive use of floor space for containers of chemicals, equipment, and/or other items restrict aisleways. Immed  
WTHR 442 Fuchs Excess flammables stored outside of approved cabinet and/or safety cans. 60 days 12/19/2002
WTHR 442 Fuchs Hoods function properly when work is set up 6" from face opening. Add a tape or paint line to any/all hoods where necessary to remind users to keep the work 6" back from sash. 60 days 1/7/2003
WTHR 442 Fuchs Excessive use of floor space for containers of chemicals, equipment, and/or other items restrict aisleways. Immed 12/19/02
WTHR 442 Fuchs Hazard assessment certification is posted but is not appropriate to this work area. Remove WELDING and any other sections not necessary. Ensure that rules are clear, appropriate and enforced. 10 days 1/24/03
WTHR 442 Fuchs Containers of chemicals are inadequately labeled. If acronyms, formulae, or abbreviations are used, post a legend/key near inside of entrance to room. See http://www.purdue.edu/rem/home/booklets/Acronyms.xls Immed 12/19/2002
WTHR 450A Fuchs Excessive use of floor space for containers of chemicals, equipment, and/or other items restrict aisleways. Immed 12/19/2002
WTHR 450A Fuchs Hoods function properly when work is set up 6" from face opening. Add a tape or paint line to any/all hoods where necessary to remind users to keep the work 6" back from sash. 60 days 1/7/03
WTHR 450D Fuchs Hoods function properly when work is set up 6" from face opening. Add a tape or paint line to any/all hoods where necessary to remind users to keep the work 6" back from sash. 60 days 1/7/03
WTHR 450D Fuchs Hazard assessment certification is posted but is not appropriate to this work area. Remove WELDING and any other sections not necessary. Ensure that rules are clear, appropriate and enforced. 10 days 1/24/03
WTHR 450E Fuchs Excessive use of floor space for containers of chemicals, equipment, and/or other items restrict aisleways. Immed

12/19/02

WTHR 450E Fuchs Hoods function properly when work is set up 6" from face opening. Add a tape or paint line to any/all hoods where necessary to remind users to keep the work 6" back from sash. 60 days 1/7/03
WTHR 454 Negishi Workplace not reasonably free of recognized hazards. Liquid in refrigerator without secure closure, parafilm is NOT allowed as only closure for liquids in refrigerators. 10 days 12/23/2002
WTHR 454 Negishi Hoods function properly when work is set up 6" from face opening. Add a tape or paint line to any/all hoods where necessary to remind users to keep the work 6" back from sash. 60 days 1/10/2003
WTHR 454 Negishi Excessive use of floor space for containers of chemicals, equipment, and/or other items restrict aisleways. (Although in general the lab and benches look very tidy.) Immed 12/14/2002
WTHR 463 Grutzner Failure to post written Hazard Assessment Certification. See http://www.purdue.edu/rem/home/forms/HazAsTmp.doc and follow instructions/directions. 5 days  
WTHR 463 Grutzner Excessive use of floor space for containers of chemicals, equipment, and/or other items that restrict aisleways. Immed  
WTHR 463 Grutzner Excess flammables stored outside of approved cabinet and/or safety cans. 60 days  
WTHR 465 Negishi Excessive use of floor space for containers of chemicals, equipment, and/or other items restrict aisleways. Immed 12/14/2002
WTHR 465 Negishi Restore housekeeping. Severe problem with many items sticking out from benchtops into walking areas. 10 days 12/20/2002
WTHR 465 Negishi Hoods function properly when work is set up 6" from face opening. Add a tape or paint line to any/all hoods where necessary to remind users to keep the work 6" back from sash. 60 days 1/10/2003
WTHR 465 Negishi Hazard assessment certification is posted but refers to a non-existent page 2. Personal protective equipment must be specific. 30 days 1/10/2003
WTHR 465 Negishi Open chemical containers not in use. Immed 12/14/2002

*Room number cited by inspectors is apparently incorrect.  Room number in red is correct.